|
|
||||||||

1Laboratory of Medical Microbiology, University of Antwerp UA, Wilrijk, Belgium 2Laboratory of Clinical Microbiology, University Hospital Antwerp UZA, Wilrijkstraat 10, B-2650 Edegem, Antwerp, Belgium 3Laboratory of Molecular Microbiology, St Elisabeth Hospital, Hilvarenbeekseweg 60, NL-5000 AS Tilburg, The Netherlands 4Health Inspectorate of the Province of Antwerp, Ministry of the Flemish Community, Antwerp, Belgium
Correspondence Kristien Dirven kristien.dirven{at}ua.ac.be
Received 28 September 2004
Accepted 22 August 2005
During an outbreak of legionellosis in Belgium, urine samples of 32 legionellosis patients were tested with three Legionella urinary antigen assays: the Biotest enzyme immunoassay (EIA) kit, the Binax EIA kit and the Binax NOW Immunochromatographic Test kit. The three tests were concomitantly compared. The test sensitivities on the first urine samples were 65.6 % for the Biotest EIA, 50.0 % for the Binax EIA and 56.3 % for the Binax NOW. Testing of a second urine sample increased the sensitivities to 71.9 %, 59.4 % and 65.6 %, respectively. The differences were not statistically significant. In outbreak settings, testing second samples from patients presenting with symptoms but initially testing negative and/or concentrating urine samples for testing might be valuable additions to the urinary antigen test to increase the sensitivities of the tests.
Present address: Center for General Practice, Building S, Room S5.40, University of Antwerp UA, Universiteitsplein 1, B-2610, Wilrijk, Belgium. Abbreviations: EIA, enzyme immunoassay; ICT, immunochromatographic test.
| INTRODUCTION |
|---|
|
|
|---|
Although fluorescent-antibody staining of sputum is a rapid detection method, its routine use is limited, mostly due to the variable and rather low sensitivity reported (2570 %) and the specificity, which greatly depends on the quality of the antibodies used (Edelstein, 1993; Fehrenbach et al., 1995; Stout & Yu, 1997; Waterer et al., 2001). Amplification of Legionella DNA by PCR (sensitivity reported 75 % to >90 %) seems promising, especially real-time PCR, but remains to be validated for the rapid diagnosis of Legionella infections (Ballard et al., 2000; Cloud et al., 2000; Templeton et al., 2003; Welti et al., 2003; Wilson et al., 2003).
During an outbreak of legionellosis after a trade fair in Kapellen, a small town in northern Belgium, we used, evaluated and compared three common urinary antigen assays, the Binax Legionella Urinary Antigen EIA, the Binax NOW Legionella Immunochromatographic Test (ICT) (Binax) and the Biotest Legionella Urine EIA (Biotest), for rapid diagnosis of visitors presenting to local hospitals with symptoms possibly suggesting legionellosis (De Schrijver et al., 2000, 2003).
| METHODS |
|---|
|
|
|---|
Urine antigen assays
Biotest Legionella EIA. The Biotest Legionella EIA is a direct sandwich assay using polyclonal rabbit immunoglobulin G as capture and detection probes for antigens of all L. pneumophila serogroups as well as for antigens of other Legionella species. The test was performed according to the manufacturer's instructions and took less then 4 h. The cut-off value for a positive result was calculated from the mean OD405 value of two negative controls plus 0.200. Urine samples with an OD405 below the cut-off value were considered negative, samples with an OD405 equal to or greater than the cut-off were considered positive. Weak positive samples in the region of the cut-off+0.200 were tested again for confirmation and were considered positive if the repeat result was above the cut-off value.
Binax Legionella EIA.
The Binax Legionella EIA is a direct sandwich assay, using polyclonal rabbit immunoglobulin G as the capture and detection probes, but specific for L. pneumophila serogroup 1 antigen. Following the manufacturer's instructions this test took less than 4 h. A sample was considered positive when the ratio of the patient urine absorbency/mean absorbency of two negative controls was
3. Urine samples with a ratio < 3 were considered negative.
Binax NOW Legionella ICT. The Binax NOW Legionella ICT is an immunochromatographic assay using rabbit anti-L. pneumophila serogroup 1 antibody as the capture probe and rabbit anti-L. pneumophila serogroup 1 antibody conjugated to colloidal gold as the detection probe. Performed according to the manufacturer's instructions the reaction could be read within 15 min, by the presence or absence of visually detectable pink-to-purple coloured lines. The appearance of both a control and a patient line corresponds to a positive result, whereas a negative assay reveals only a control line. Absence of any line indicates an invalid assay, requiring the test to be repeated.
Second urine samples.
From 17 confirmed patients who were tested with all three urinary antigen assays, a second urine sample was available for additional testing with the three assays. Ten of these had symptoms suggesting legionellosis, but tested negative in their first sample. They were all retested within 2 weeks of the end of the fair. Patients whose health deteriorated after a first urine sample had been collected were also asked for a second urine specimen.
Statistical analysis.
Data were analysed using the Epi-Info-software, version 6.0.
| RESULTS |
|---|
|
|
|---|
Considering only the first urine specimens, the Biotest EIA detected 21, the Binax EIA 16 and the Binax NOW ICT 18 cases. The differences are not statistically significant (Biotest EIA vs Binax EIA, P = 0.21; Biotest EIA vs Binax NOW ICT, P = 0.44; Binax EIA vs Binax NOW ICT, P = 0.61) (Table 1).
|
Testing of second urine specimens resulted in two more cases being detected (after 3 and 9 days) with the Biotest EIA and three more cases (one after 2 days, two after 3 days) with the Binax EIA as well as the Binax NOW ICT. When these test results were included, the sensitivity of the Biotest EIA, the Binax EIA and the Binax NOW ICT improved slightly. The observed differences were not statistically significant (Biotest EIA vs Binax EIA, P = 0.29; Biotest EIA vs Binax NOW ICT, P = 0.59; Binax EIA vs Binax NOW ICT, P = 0.60) (Table 1).
Discrepancies between the three urinary antigen assays
For six of the 32 patients (18.8 %) discrepancies between the three urinary antigen assays were observed after testing the first urine samples.
One patient tested positive with the Binax NOW ICT in the first urine sample, but remained negative with the other two tests. This patient tested negative with all three tests on the second urine sample. Another patient tested positive for Biotest EIA and Binax NOW ICT in the first urine sample, but negative for Binax EIA. No second urine sample was available for further testing. The remaining four patients tested positive with the Biotest EIA, but negative with the other two tests. For one patient no second urine sample was available. One patient tested positive with all three tests on the second urine sample. Two patients tested negative with all the three tests on the second urine sample. One of these patients later showed seroconversion.
| DISCUSSION |
|---|
|
|
|---|
When the results of the first urine samples of the confirmed patients were evaluated, the sensitivities of the Biotest EIA (65.6 %), the Binax EIA (50.0 %) and the Binax NOW ICT (56.3 %) were surprisingly low. The differences between the sensitivities of the three tests were not significant. These sensitivities were similar to those seen in the 1999 outbreak in The Netherlands (6972 %) (Yzerman et al., 2002) and in a study by Domínguez et al. (1999) (55.5 % for the Binax NOW ICT), but were different to those reported elsewhere (Biotest and Binax EIA, 71100 %; Binax NOW, 78.697 %; Benson et al., 2000; Harrison et al., 1998; Helbig et al., 2001, 2003; Horn, 2001; Kashuba & Ballow, 1996; Kazandjian et al., 1997; Ruf et al., 1990).
As the three assays primarily detect L. pneumophila serogroup 1 infections and this was the causative agent in the Kapellen epidemic, the low test sensitivities cannot be explained by the possibility of missing cases of Legionnaires disease caused by other serogroups and species. They could, however, be explained by a couple of other factors. It was suggested by Wever et al. (2000) and Yzerman et al. (2002) that there might be an association between the severity of the disease and test sensitivity. Yzerman et al. (2002) demonstrated that the test sensitivitiy for patients with mild disease is much lower than for severe cases (42 % compared to 75 %), which might explain our findings, as active case-finding probably brought more patients with mild legionellosis to the hospital. Another factor worth considering is the variation of antigen excretion in time. Most patients were detected by the examination of their first urine specimen, collected 2 to 15 days after the end of the fair. It is, however, possible that some first urine samples were collected before antigen appeared in the urine, usually within days after onset of illness, as cases were actively sought and many patients presented early. This might explain some discrepancies observed between negative first and positive second urine specimens. Domínguez et al. (1996, 1998) showed that the concentration of urine is important as well. Performing urinary antigen assays on concentrated urine samples improves sensitivity (Yzerman et al., 2002). Our limited experience seems to add to this, as four of six previously negative patients for whom we could concentrate urine gave positive results for the retest using the concentrated sample with the Biotest NOW ICT (increasing the sensitivity of the test by 12.5 %).
Examination of the second urine specimens increased the sensitivity considerably, to 71.9 %, 59.4 % and 65.6 %, respectively, for the Biotest EIA, the Binax EIA and the Binax NOW ICT, but the results remain at the low side of the range reported in the literature. The differences between the three tests were not significant. Again the factors mentioned above might explain the discrepancies observed between first and second urine samples.
Whenever an outbreak of legionellosis is suspected, we strongly recommend the examination of concentrated urine specimens and/or the testing of a second specimen a few days later for those patients presenting with symptoms but initially testing negative.
The results of both the Biotest EIA and the Binax EIA can be available within 4 h of receiving the urine specimen, whereas the Binax NOW ICT produces results within 1520 min. Therefore, the Binax NOW ICT can be valuable when the time factor is important. Furthermore, it can be a valuable asset in those laboratories that are not familiar with the Legionella EIA kits as it is very easy to perform and does not require any additional equipment. On the other hand, the EIA kits may be preferred for reference laboratories, laboratories with a large demand for Legionella testing and during outbreaks, since larger numbers of specimens can be tested in a single batch, allowing a significant reduction in costs.
We conclude that, as long as PCR is not more widely evaluated and validated, the urinary antigen tests continue to play a very important role in the rapid diagnosis of outbreak-associated cases of legionellosis.
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
| REFERENCES |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| INT J SYST EVOL MICROBIOL | J MED MICROBIOL | MICROBIOLOGY | J GEN VIROL | ALL SGM JOURNALS |