|
|
||||||||
1Specialist and Reference Microbiology Division, 61 Colindale Avenue, Colindale, London NW9 5HT, UK 2Chelmsford Microbiology Laboratory, Chelmsford, UK 3Broomfield Hospital, Chelmsford, UK
Correspondence Stephanie A. Chisholm Stephanie.Chisholm{at}hpa.org.uk
Received October 8, 2003
Accepted March 15, 2004
Stool antigen-testing allows non-invasive detection of Helicobacter pylori that is indicative of active infection. Three commercial kits are currently marketed in the UK for stool antigen-testing. The aim of this study was to conduct a comparative evaluation of the performances of each of these tests, compared with culture and histological examination of gastric biopsies, for pre-treatment diagnosis of infection in an adult dyspeptic population in south-east England. Examination of 112 stool samples by the Premier Platinum HpSA ELISA (Meridian Diagnostics) and by the Amplified IDEIA HpStAR ELISA (DakoCytomation) kits demonstrated that the latter was more sensitive (81.3 versus 93.8 %, respectively) and specific (91.7 versus 100.0 %, respectively). Additionally, the IDEIA HpStAR was easier to interpret, with OD readings of positive and negative results being far from the recommended cut-off, whereas equivocal results that were generated by the HpSA kit were difficult to interpret. Additional testing of 87 of the 112 stools by the ImmunoCard STAT! HpSA kit (Meridian Diagnostics) demonstrated that this test was easier to perform than ELISA and was more sensitive than the HpSA kit but, compared with the IDEIA HpStAR kit, the ImmunoCard test was less sensitive (87.8 versus 95.9 %, respectively) and specific (89.4 versus 100.0 %, respectively). Furthermore, the ImmunoCard test generated weakly positive results, correlating with lower OD readings for both ELISA kits, that were difficult to interpret. The Amplified IDEIA HpStAR kit is therefore the most sensitive and specific of the three tests that are available for pre-treatment, non-invasive detection of H. pylori in stool samples in an English adult dyspeptic population.
| INTRODUCTION |
|---|
|
|
|---|
Two commercial ELISAs, the Premier Platinum HpSA kit (Meridian Diagnostics) and the Amplified IDEIA HpStAR kit (formerly the FemtoLab Cnx H. pylori kit) (DakoCytomation), are currently marketed in the UK for stool antigen-testing. Since the original evaluative study of the Premier Platinum HpSA kit (Vaira et al., 2000), numerous studies have shown it to be highly sensitive and specific when applied to adult (Odaka et al., 2002; Vaira et al., 2000, 2002) and paediatric (Oderda et al., 2000; Konstantopoulos et al., 2001) populations for testing pre-treatment (Makristathis et al., 2000; Oderda et al., 2000; Gisbert & Pajares, 2001; Konstantopoulos et al., 2001) and post-therapy (Makristathis et al., 2000; Konstantopoulos et al., 2001; Odaka et al., 2002; Vaira et al., 2002). The observations of these individual studies were confirmed by a recent meta-analysis of 68 studies, which reported an overall sensitivity and specificity of 92.4 and 91.9 %, respectively, for diagnosing infection in untreated patients (Gisbert & Pajares, 2001) and of 88.3 and 92 %, respectively, for patient follow-up at
4 weeks to assess eradication outcome (Gisbert & Pajares, 2001). Results from individual studies suggested that the sensitivity and specificity of this test may vary according to geographical locations (Gisbert & Pajares, 2001), demonstrating the need for local evaluation of kit performance prior to routine implementation. Whilst one study has examined the performance of the HpSA kit in a paediatric patient group in south-west Scotland (Shepherd et al., 2000), no other evaluative studies have been conducted in the UK. Although few evaluative studies of the IDEIA HpStAR kit have been reported, evidence to date suggests that this kit is marginally more sensitive than the HpSA kit for primary diagnosis of paediatric infection (Makristathis et al., 2000; Koletzko et al., 2003) and for post-treatment follow-up of such patients (Makristathis et al., 2000). Similar results were reported in a study that investigated 148 adult dyspeptic patients after eradication therapy (Leodolter et al., 2002). Additionally, the ImmunoCard STAT! HpSA test (Meridian Diagnostics) is now available in the UK, but no evaluative studies of its performance have yet been reported.
This study evaluated both ELISA kits (Premier Platinum HpSA and Amplified IDEIA HpStAR) and the ImmunoCard STAT! HpSA test, a lateral-flow immunochromatography test, to compare relative performances when applied to stools from dyspeptic adult patients in south-east England.
| METHODS |
|---|
|
|
|---|
Local ethical approval was obtained to collect stool samples from patients at the time of endoscopy; these were provided with the patients informed consent. Stools were stored immediately at 20 °C until they were analysed. All 112 stools were tested for H. pylori antigen by the Premier Platinum HpSA and by the Amplified IDEIA HpStAR ELISA kits, as described below. Additional testing by the ImmunoCard STAT! HpSA kit was performed on 87/112 stools, 49 of which were from H. pylori-positive patients.
Detection of H. pylori stool antigen by the Premier Platinum HpSA kit.
H. pylori antigen was detected from stools by the Premier Platinum HpSA ELISA (Launch Diagnostics) by following the protocol of the manufacturer (Meridian Diagnostics). Briefly, a 56 mm-sized stool sample (or 100 µl, if liquid) was emulsified in 200 µl sample diluent and mixed by vortexing. Diluted stool (50 µl) was added to an anti-H. pylori antibody-coated well, along with one drop of peroxidase-conjugated antibody. Plates were incubated at room temperature for 1 h and unbound material was removed by five manual washes with the buffer provided. Bound antigenantibody complexes were detected by a colour change, following incubation in the dark with enzyme substrate at room temperature for 10 min. Reactions were stopped by the addition of stop solution and spectrophotometric absorbances were measured at 450 and 630 nm, within 30 min, using a Labsystems Multiskan RC version 6.0 (Labsystems Oy). According to the manufacturer's guidelines, samples were defined as either negative, equivocal or positive on the basis of OD450/OD630 readings of <0.100, 0.1000.120 and
0.120, respectively.
Detection of H. pylori stool antigen by the Amplified IDEIA HpSTAR kit.
H. pylori-specific antigen was detected by the Amplified IDEIA HpSTAR ELISA kit, following the procedure provided by the manufacturer (DakoCytomation). Approximately 0.1 g stool was emulsified in 500 µl sample diluent, vortexed and then centrifuged (5000 r.p.m. for 5 min). Supernatant (50 µl) was added to antibody-coated wells, along with 50 µl peroxidase-conjugated anti-H. pylori antibody; this was incubated at ambient temperature for 1 h. Wells were washed by using a Well Wash 4 automated plate-washer (Denley, Life Sciences International UK) and remaining bound complexes were detected by incubation at ambient temperature for 10 min in the dark with 100 µl substrate. Following the addition of 100 µl stop solution, results were recorded as described for the HpSA kit. Samples were defined as either negative or positive on the basis of OD450/OD630 readings of <0.150 and
0.150, respectively, as indicated by the manufacturer.
Detection of H. pylori stool antigen by the ImmunoCard STAT! HpSA kit.
Stool samples were tested for H. pylori antigen by the ImmunoCard STAT! HpSA kit (Launch Diagnostics) by following the protocol of the manufacturer (Meridian Diagnostics). Briefly, a 56 mm-sized stool sample was added to vials that contained 1 ml sample diluent and emulsified by vortexing for 15 s. The tip of the vial was snapped off and four drops were added to the sample port of the test cassette. The test was read after exactly 5 min incubation at ambient temperature. Tests were interpreted as negative if there was a blue line in the control (C) window only, and as positive if there was any evidence of an additional pink line in the test (T) window. Results were defined as positive if the pink band was clearly visible and weakly positive if it was very faint.
| RESULTS AND DISCUSSION |
|---|
|
|
|---|
|
|
The quality of results generated by the two kits, indicated by OD450/OD630 readings, also varied (Fig. 1). The mean OD450/OD630 values recorded for HpSA-positive and -negative specimens were 0.447 ± 0.137 units above and 0.059 ± 0.008 units below the kit's cut-off value for equivocal results of 0.10 OD units, respectively (Fig. 1). The mean OD450/OD630 values of the IDEIA HpStAR kit for positive and negative results were 1.930 ± 0.344 units above and 0.113 ± 0.006 units below the kit's recommended cut-off level (0.15 OD units) (Fig. 1). Comparison by an unpaired t-test demonstrated that the OD450/OD630 values of the IDEIA HpStAR kit were significantly further above or below the defined cut-off than those of the HpSA kit (P < 0.0001) and were therefore easier to interpret.
The observed variations in performance may relate to the different capture antibodies used in each assay. The monoclonal capture antibodies in the HpStAR kit would allow highly specific binding of H. pylori antigen. In contrast, the polyclonal antibodies of the HpSA kit, whilst potentially more sensitive, could theoretically bind non-specifically to a wider range of faecal antigens, leading to higher background OD levels. Use of polyclonal antibodies could also lead to variation in assay performances between kit batches. Additionally, preliminary centrifugation of diluted stool in the HpStAR kit removes insoluble faecal material, prior to incubation of samples with capture and conjugate antibodies. This may improve accessibility and diffusion of soluble H. pylori antigen and, therefore, facilitate formation of specific antigenantibody complexes, leading to higher OD levels for positive results. Removal of solid particles also allowed washing steps to be automated and hence standardized between runs. In contrast, the HpSA kit protocol does not include a centrifugation step and both insoluble and soluble faecal material is incubated with the capture and conjugate antibodies, which may lower the rate of specific complex formation and, ultimately, OD readings, while increasing non-specific binding and background OD levels. Automated well-washing was not possible, because of the potential for equipment blockage, and efficiency of manual washing may have varied between runs. Although care was taken to ensure adequate washing of wells, the amount and nature of insoluble material varied between specimens. Levels of residual material may also have contributed to the high non-specific, background OD levels that were observed in some samples.
Our results suggest that of the two commercially available, ELISA-based kits for H. pylori stool antigen detection, the IDEIA HpStAR kit is the most accurate for pre-treatment H. pylori diagnosis in adult dyspeptic patients in south-east England. Few studies have compared the relative performances of these two kits (Makristathis et al., 2000; Leodolter et al., 2002; Koletzko et al., 2003). The findings of this study of an adult dyspeptic population were in agreement with those of an earlier study that demonstrated marginally higher sensitivity (98.0 %) of the IDEIA HpStAR (=FemtoLab) kit for primary diagnosis of H. pylori infection in 49 paediatric patients when compared with the HpSA kit (93.8 %) (Makristathis et al., 2000). Additionally, that study demonstrated that both kits allowed patient follow-up post-therapy. A subsequent study of 148 adult patients also reported marginally higher sensitivity of the HpStAR kit for evaluating success of eradication therapy, although the kit cut-off OD was lowered to 0.90 to achieve this (Leodolter et al., 2002). None of the patients examined in our study had received specific eradication therapy, but validation of each kit for adult patient follow-up is needed so that appropriate cut-offs can be established. Likewise, further studies are essential to fully evaluate the performances of these kits in paediatric patients from the UK.
Comparison of ImmunoCard STAT! HpSA test and ELISA kits
For the 87 stool specimens analysed by all three tests, specific antigen was detected by the HpSA and IDEIA HpStAR ELISA kits in, respectively, 39/49 and 47/49 H. pylori-positive patients and in 2/38 and 0/38 H. pylori-negative patients. Positive results were obtained for the ImmunoCard test in 43/49 positive patients and in 4/38 negative patients. Respective sensitivities for the HpSA, the IDEIA HpStAR and the ImmunoCard tests were thus 79.6, 95.9 and 87.8 %, while test specificities were 94.7, 100.0 and 89.4 %, respectively. However, if five equivocal results generated by the HpSA ELISA were considered negative, sensitivity of that test was 71.4 %, while specificity was 97.4 %.
Positive results by the ImmunoCard STAT! HpSA test could be subdivided into weak, moderate or strong, based on the intensity of the pink band. Stool specimens that were strongly or moderately positive generated mean OD450/OD630 readings of 0.527 ± 0.178 for the HpSA kit and 2.72 ± 0.333 for the HpStAR kit. In contrast, mean OD450/OD630 readings for stools that were weakly positive or falsely negative were significantly lower for both the HpSA kit (0.146 ± 0.057, P = 0.0009) and the HpStAR kit (0.774 ± 0.335, P < 0.0001). Further illustration of this correlation is provided by the observation that stools that were falsely negative or equivocal by ELISA were also weakly positive or falsely negative by ImmunoCard HpSA test (Table 2). Weakly positive ImmunoCard results could be difficult to interpret, particularly as faint pink bands also could be observed in negative stools beyond the recommended 5 min incubation. High sensitivity and specificities have been reported for similar ImmunoCard tests that are available for the detection of enteric pathogens Escherichia coli O157 (Mackenzie et al., 2000) and Giardia lamblia and Cryptosporidium parvum (Garcia et al., 2003). Our results suggest that the ImmunoCard HpSA test was more sensitive, but less specific, than the HpSA ELISA and allows accurate detection of H. pylori from stools with higher levels of antigen. However, weakly positive results were defined subjectively and, as for the equivocal results of HpSA ELISA, would be difficult to interpret in a clinical setting.
|
Both ELISA-based kits required inclusion of a positive and a negative control in each run. It was therefore most economical to batch-test multiple stool specimens in a single experiment. In contrast, the immunochromatographical ImmunoCard STAT! HpSA test is more suitable for testing single or small numbers of stools. This test uses mAbs to capture specific antigen and results can be obtained in 510 minutes, compared to 23 h with ELISA-based methods. The speed and simplicity of this test are clearly advantageous compared to ELISA, particularly in a laboratory that examines low numbers of stool specimens.
In conclusion, our study has shown that whilst the ImmunoCard STAT! HpSA test is more convenient to use than either ELISA kit, the higher sensitivity and specificity of the IDEIA HpStAR ELISA kit support its use for routine pre-treatment diagnosis of H. pylori infection in an adult population in south-east England.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Krausse, G. Muller, and M. Doniec Evaluation of a Rapid New Stool Antigen Test for Diagnosis of Helicobacter pylori Infection in Adult Patients J. Clin. Microbiol., June 1, 2008; 46(6): 2062 - 2065. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Megraud and P. Lehours Helicobacter pylori Detection and Antimicrobial Susceptibility Testing Clin. Microbiol. Rev., April 1, 2007; 20(2): 280 - 322. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| INT J SYST EVOL MICROBIOL | J MED MICROBIOL | MICROBIOLOGY | J GEN VIROL | ALL SGM JOURNALS |