|
|
||||||||
Departments of Medicine1 and Microbiology2, Queen Mary Hospital, The University of Hong Kong, Hong Kong, China
Correspondence Tak-Mao Chan dtmchan{at}hku.hk
Received September 11, 2002
Accepted January 20, 2003
| Abstract |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
There is increasing evidence that measuring the level of HBV DNA in serum is useful in monitoring the efficacy of antiviral therapy, detecting the occurrence of drug-resistant mutants and detecting relapse after discontinuing antiviral therapy (Chan et al., 1998, 2002; Lai et al., 1998; Omata, 1998). Assays to quantify HBV DNA that are currently available commercially are based on either direct nucleic acid hybridization (Hendricks et al., 1995; Ho & Chan, 2000; Lai et al., 1999) or competitive PCR (Kessler et al., 1998; Noborg et al., 1999). Amongst the former, we and others have reported the superiority of the Hybrid Capture II HBV DNA test (HCII) (Digene Corp.) over other non-PCR-based assays in terms of accuracy and sensitivity (Ho et al., 1999; Ho & Chan, 2000; Niesters et al., 2000). To date, although PCR-based assays have increased sensitivity, the technical complexity, susceptibility to contamination and more variable reproducibility of results with these assays have hindered their application in clinical practice (Kaneko et al., 1990; Quint et al., 1995).
Recent advances in PCR instrumentation and the capacity for simultaneous fluorimetric detection of PCR amplification products have enabled real-time kinetic analysis of the latter (Higuchi et al., 1992; Holland et al., 1991). Since amplification, measurement and quantification of PCR product occur simultaneously in the same closed reaction vessel, the need for post-PCR manipulations is obviated and the risk of PCR product carry-over contamination is minimized. Real-time PCR technology has been used to measure HBV DNA in serum using the TaqMan probe in the ABI Prism 7700 sequence detector (Perkin Elmer Applied Biosystems) (Abe et al., 1999; Loeb et al., 2000; Pas et al., 2000; Weinberger et al., 2000). In the present study, we have developed a rapid, specific and highly sensitive real-time PCR assay for HBV DNA based on the fluorescence resonance energy transfer (FRET) principle, with a pair of fluorescent hybridization probes and the LightCycler system (LC-PCR) (Roche Diagnostics). The accuracy, intra-assay and inter-assay variability and the detection range for this novel assay were examined, together with validation using EUROHEP HBV DNA standards. Also, the clinical performance of the LC-PCR was compared with that of the commonly used HCII test.
| METHODS |
|---|
|
|
|---|
HCII assay.
The HCII (`standard') assay (Digene Corp.) quantified HBV DNA by solution hybridization, immunocapture and chemiluminescent signal detection. The assay protocol followed the manufacturer's instructions. Briefly, 30 µl denaturing reagent was added to each microplate well containing 30 µl of test samples or HBV DNA standards (06000 pg ml-1). The plate was incubated at 65 °C for 30 min to allow lysis of HBV and DNA denaturation. RNADNA hybridization was achieved by adding 30 µl RNA probe (specific for HBV ad and ay strains) to each well and incubating at 65 °C for 60 min. An aliquot (75 µl) of the hybrid-containing solution was then transferred into RNADNA capture wells and shaken (Thermolyne Maxi-Mix III) at 1100 r.p.m. at room temperature for 60 min. The solution in the wells was then removed by aspiration. An aliquot (75 µl) of alkaline phosphatase-conjugated antibodies to RNADNA hybrids was added to each well and incubated at room temperature for 30 min. After six washings, 75 µl chemiluminescent substrate was added and light emission after 15 min was measured with a chemiluminometer (DML 2000 Luminometer; Digene Corp.). Results were expressed in pg ml-1 according to the plot of standards. The sensitivity according to the manufacturer was 0.5 pg or 1.42 x 105 copies ml-1. We have previously reported an evaluation of this assay (Ho et al., 1999).
LC-PCR assay.
Nucleic acid was extracted from 200 µl serum and Digene Calibrator 5 using the QIAamp DNA Blood Mini kit (Qiagen) according to the manufacturer's protocol. DNA was eluted from the QIAamp spin column with 50 µl distilled water, which served as the template for the LC-PCR assay. This DNA preparation had a concentration factor of 4 and a yield of 83 ± 3 % compared with the original amount of DNA in the sample. Calibrator 5 from the HCII assay, which contained 1.7 x 109 copies ml-1 (6000 pg ml-1) HBV plasmid DNA, was used to prepare the calibration curve for the LC-PCR assay, with a range from 101 to 108 HBV DNA copies per reaction. The Qiagen DNA extraction step was performed on this Calibrator 5 standard as on the other samples to ensure parity of nucleic acid yield. In addition, this step removed the sodium azide used as a preservative in the Calibrator 5 standard, which might affect the PCR.
PCR was performed using a Roche Diagnostics LightCycler and the fluorescent signal of the PCR products after each cycle was monitored continuously by the FRET principle with a pair of fluorescent hybridization probes (Espy et al., 2000; Ghosh et al., 1994). We used a set of primers for the HBV core region (BcP1, 5'-ACCACC AAATGCCCCTAT-3'; BcP2, 5'-TTCTGCGACGCGGCGA-3'), according to published nested PCR protocols (Chung et al., 1993; Ho et al., 1999), which yielded a 130 bp product. The donor fluorescein probe (HBVcD, 5'-GAGTTCTTCTTCTAGGGGACCTGCfluorescein-3') and the acceptor LightCycler-Red 640 (LCRed) probe (HBVcA, 5'-LCRedTCGTCGTCTAACAACAGTAGTTTCCG-phosphate-3') directed to the 130 bp product were designed according to the manufacturer's guidelines (Roche). Aliquots of 10 µl of the extracted nucleic acid template and 10 µl of a reaction mixture containing 2 µl 10 x LightCycler FastStart DNA Master hybridization probes (Roche), 1 µM BcP1, 1 µM BcP2, 0.3 µM HBVcD donor probe, 0.3 µM HBVcA acceptor probe and 5 mM MgCl2 were added to each capillary. All capillaries were sealed and then centrifuged briefly before amplification in the LightCycler. The PCR protocol consisted of an initial step at 95 °C for 10 min for activation of the FastStart Taq DNA polymerase, followed by 45 cycles of 0 s (hold time on reaching temperature) denaturation at 95 °C, 12 s annealing at 60 °C and 12 s extension at 72 °C. The temperature transition rate was 20 °C s-1 for all steps. Fluorescence data were acquired once each cycle at the end of the annealing phase with detection channel setting at F2/F1.
For data analysis, baseline adjustment was carried out in the proportional mode and fluorescence curve analyses were carried out in the fit points mode of the LightCycler software. Positive results obtained in the LC-PCR assay using fluorescent hybridization probes were characterized by a sigmoid curve, showing an initial, rapid, exponential increase in fluorescence signal followed by a plateau. Negative reactions did not show any increase in fluorescence signal. A standard curve was created automatically with the LightCycler software in each run by plotting the threshold cycle number against the copy numbers of each standard and quantification of HBV DNA for unknown samples was inferred from the regression line.
Statistics.
Determination of the inter- and intra-assay coefficients of variation (CV) of the LC-PCR in testing the EUROHEP standards at different dilutions and the xy scatter plot, Spearman's correlation (r) and logarithmic transformation of the readings for comparison of the HCII and LC-PCR assays were done by using the SPSS software (version 8.0, SPSS Inc.). Comparisons between HBeAg+ and anti-HBe+ subjects were by
2 and MannWhitney tests.
| RESULTS |
|---|
|
|
|---|
|
To determine the accuracy and CV of the LC-PCR assay at different HBV DNA concentrations, EUROHEP standards from a single extraction were tested in three separate runs, both undiluted and in sequential 10-fold dilutions. An almost linear results curve against the theoretical values was obtained over the complete range of 8 orders of magnitude for both ad and ay serotypes, with intra-assay CV of 6 % and inter-assay CV of 16 % (Fig. 2). All samples with ad subtype concentration
270 copies ml-1 or ay subtype concentration
260 copies ml-1 tested positive, confirming the sensitivity limit of 250 copies ml-1 indicated by the standard curve (Fig. 2).
|
To examine the clinical performance of the LC-PCR assay, HBV DNA concentrations were measured in serum samples from 120 HBsAg+ patients and 45 HBsAg- healthy controls and the results were compared with those determined using the HCII assay. The seroprevalence for HBV DNA among the HBsAg+ patients was 114/120 (95 %) by LC-PCR and 67/120 (56 %) by HCII (P < 0.01) (Table 1). All 67 HCII-positive samples yielded positive results with the LC-PCR assay, and the log-transformed HBV DNA concentrations obtained with the two assays demonstrated a good correlation (r = 0.902; P < 0.001). The 47 discordant samples from HBsAg+ patients that tested negative by HCII (i.e. < 1.42 x 105 copies ml-1) but positive by LC-PCR contained HBV DNA at low concentrations, down to 265 copies ml-1 (Fig. 3). All 45 samples from healthy controls tested negative by both assays, thereby yielding specificity rates of 100 %.
|
|
Of the 120 HBsAg+ patients, 54 (45 %) were positive for HBeAg and 66 (55 %) were positive for anti-HBe, and the prevalence of HBV DNA by LC-PCR was similar in the two groups (100 and 95 %, respectively; not significant). In contrast, the HCII assay failed to detect HBV DNA in a considerable number of anti-HBe+ patients (89 vs 29 %; P < 0.001) (Table 1). The HBV DNA concentration, as determined by LC-PCR, was significantly higher in HBeAg+ patients compared with those positive for anti-HBe (median 1.5 x 107 vs 4.6 x 104 copies ml-1, range 11006.6 x 109 and < 2501.8 x 109; P < 0.05) (Fig. 4).
|
| DISCUSSION |
|---|
|
|
|---|
We have designed the sequences of the donor (HBVcD) and acceptor (HBVcA) hybridization probes on the basis of a set of primers (BcP1 and BcP2) that we have used for more than 8 years in our in-house HBV nested PCR assay (Chung et al., 1993; Ho et al., 1999). The upstream probe is labelled at the 3'-end with fluorescein and the downstream probe at its 5'-end with LCRed. As the PCR products are made, FRET from fluorescein (donor) to LCRed (acceptor) results in an increase in LCRed signal. The LCRed signal is determined by the amount of product at the end of each PCR cycle, and the first significant increase in signal during the exponential phase (i.e. the threshold cycle) correlates with the initial amount of target DNA template. Our probes have been designed to anneal one base apart on the target 130 bp amplicon in order to achieve maximum FRET signal production.
The use of the QIAamp DNA Blood kit for nucleic acid extraction minimizes contamination and ensures reproducibility of yield, both important properties of quantitative assays. The LightCycler FastStart DNA Master hybridization probe also contributes to the accuracy of the LC-PCR assay, since the constituent polymerase (a modified form of thermostable recombinant Taq DNA polymerase) in this ready-to-use hot-start reaction mixture is only activated after heating at 95 °C for 10 min has removed the blocking groups.
The two EUROHEP reference standards have been used for standardization of HBV DNA test kits and in quality-control trials (Heermann et al., 1999; Niesters et al., 2000). Although it is theoretically better to use a biological standard, such as the HBV genome, rather than a plasmid calibrator, the limited supply of the EUROHEP references presents a major hindrance to their routine use in assays for clinical service. Therefore, we have used the readily available calibrator of the Digene assay as a standard in the LC-PCR assay and have validated the assay with the EUROHEP standards. Our results demonstrated linearity over the entire 8-log10 detection range, with low CV for both EUROHEP ad and ay serotypes. Evaluation of clinical samples from HBsAg+ patients and HBsAg- controls showed that the LC-PCR assay was highly specific. Its results correlated well with their HCII counterparts, thereby facilitating direct comparison of data obtained with different assays. Of particular significance is the more than 500-fold increase in sensitivity with the LC-PCR assay compared with the standard HCII test, with the ability to detect HBV DNA at levels as low as 250 copies ml-1. This presents a distinct clinical advantage in patients receiving antiviral medications, which helps to ensure adequate suppression of virus replication, in order to detect relapse early or to monitor for the emergence of drug-resistant HBV mutants. In this context, there are recent reports that have demonstrated a relationship between virus load and the propensity to develop drug-resistant mutants (Puchhammer-Stockl et al., 2000). Besides quantification, another application of fluorimetric real-time PCR assay with the LightCycler has been the detection of emergence of drug-resistant HBV mutants (YMDD motifs) in patients during lamivudine therapy (Cane et al., 1999; Whalley et al., 2001).
We have shown that the level of HBV DNA is generally higher in HBeAg+ subjects compared with those who are positive for anti-HBe and that the improved sensitivity of LC-PCR offers a definite advantage in the detection of the relatively lower levels of viraemia in the latter group. Nevertheless, there is marked individual variation, and the virus load can be high in some anti-HBe+ patients. In this regard, five anti-HBe+ patients had HBV DNA levels higher than the median level in the HBeAg+ group. Two of the five patients were renal transplant recipients and the other three were non-immunosuppressed patients with persistently abnormal liver biochemistry. These results suggest that HBeAg/anti-HBe status is not a reliable surrogate marker for virus replication or disease activity (Chan et al., 2002).
Our results demonstrate that the LC-PCR assay has a broad quantification range from 250 to 2.5 x 109 copies ml-1 and low intra-assay and inter-assay CVs, similar to assays using the TaqMan probe and ABI PRISM (Abe et al., 1999; Loeb et al., 2000; Pas et al., 2000; Weinberger et al., 2000). Other LightCycler-based quantitative HBV DNA assays have used the standard from the Chiron HBV DNA Quantiplex assay and offer detection ranges from 10 to 1010 copies per reaction (Jardi et al., 2001; Paraskevis et al., 2002). Paraskevis et al. (2002) compared home-made HBV DNA plasmid and pooled patient sera with high HBV DNA levels as standards and concluded that the two yielded very similar results. The ABI PRISM 7700 detector has the advantage of simultaneous analysis of up to 96 samples, compared with 32 samples with the LightCycler. On the other hand, the turn-around time of the LC-PCR assay is less than 2.5 h, compared with 4 h required for other real-time PCR assays. The reagents of both real-time PCR methods are less expensive than those of the commercial HBV DNA kits. Hence, the presently reported LC-PCR assay represents another option amongst the choices for real-time PCR assays; users can therefore select according to their particular requirements.
In conclusion, this LC-PCR assay permits sensitive and accurate quantification of serum HBV DNA over a wide range within 2.5 h and has potential for both research and clinical use, in view of the increasing number of patients being monitored for changes in virus load.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. J. Espy, J. R. Uhl, L. M. Sloan, S. P. Buckwalter, M. F. Jones, E. A. Vetter, J. D. C. Yao, N. L. Wengenack, J. E. Rosenblatt, F. R. Cockerill III, et al. Real-Time PCR in Clinical Microbiology: Applications for Routine Laboratory Testing Clin. Microbiol. Rev., January 1, 2006; 19(1): 165 - 256. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lindh and C. Hannoun Dynamic Range and Reproducibility of Hepatitis B Virus (HBV) DNA Detection and Quantification by Cobas Taqman HBV, a Real-Time Semiautomated Assay J. Clin. Microbiol., August 1, 2005; 43(8): 4251 - 4254. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Rouet, D. K. Ekouevi, M.-L. Chaix, M. Burgard, A. Inwoley, T. D. Tony, C. Danel, X. Anglaret, V. Leroy, P. Msellati, et al. Transfer and Evaluation of an Automated, Low-Cost Real-Time Reverse Transcription-PCR Test for Diagnosis and Monitoring of Human Immunodeficiency Virus Type 1 Infection in a West African Resource-Limited Setting J. Clin. Microbiol., June 1, 2005; 43(6): 2709 - 2717. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Q. Konnick, M. Erali, E. R. Ashwood, and D. R. Hillyard Evaluation of the COBAS Amplicor HBV Monitor Assay and Comparison with the Ultrasensitive HBV Hybrid Capture 2 Assay for Quantification of Hepatitis B Virus DNA J. Clin. Microbiol., February 1, 2005; 43(2): 596 - 603. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Stelzl, Z. Muller, E. Marth, and H. H. Kessler Rapid Quantification of Hepatitis B Virus DNA by Automated Sample Preparation and Real-Time PCR J. Clin. Microbiol., June 1, 2004; 42(6): 2445 - 2449. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Ismail, G. E. Fish, and M. B. Smith Laboratory Evaluation of a Fully Automated Chemiluminescence Immunoassay for Rapid Detection of HBsAg, Antibodies to HBsAg, and Antibodies to Hepatitis C Virus J. Clin. Microbiol., February 1, 2004; 42(2): 610 - 617. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| INT J SYST EVOL MICROBIOL | J MED MICROBIOL | MICROBIOLOGY | J GEN VIROL | ALL SGM JOURNALS |